Introduction |
|
|
This page provides a short tutorial leading through some steps that are required for obtaining U-Pb zircon ages
using the UCLA ims1270 ion microprobe: (1) Oxygen flooding (2) Energy scan and offset (3) Centering the ion image in the field aperture (4) Mass calibration (5) Presputtering and data acquisition (6) Interelement calibration and precision |
|
(1) Oxygen flooding
|
|
| Pb isotope intensity of zircon during O- ion bombardement is significantly enhanced if the sample surface is saturated with regard to oxygen. During zircon analysis, oxygen gas is leaked through a valve into the ims1270 sample chamber. Stability.vi (right) shows the ~50% increase in 206Pb counts measured on AS3 zircon that was observed while gradually opening the leak valve up to a final pressure of 2.5 10-5 Torr. Note: in order to open EP10, EP11 or to vent the door, the leak valve has to be closed (1.5 turns clockwise). | ![]() |
(2) Energy Scan and
Offset
|
|
| At the beginning of an analytical session or when analytical parameters such as primary beam intensity or spot size were changed, energy offsets for all measured ion species must be determined. Pb+, U+, and UO+ have characteristic energy distributions (see Figure) and therefore optimizing intensities and finding a region in the energy distribution sufficiently stable to minimize interelement fractionation is a prerequisite for analysis. The difference between the nominal extraction potential (= 10 kV, equivalent to 0 V offset) and the maximum of the observed energy distribution measured on a standard zircon determines the offset voltages to be applied during analysis. For the example to the right, offset values for 204Pb+ (and other Pb-species) are: +13V, for U+ (and Th+) -10V, and for UO+ +10. These values are to be entered in the condition file (offset in "analysis_species_table.vi"). | ![]() |
(3)
Centering the ion image in the field aperature
|
|
|
Centering the secondary ion image in the center of the field aperture (FA) is necessary for each new analysis spot because primary beam
paths and conditions of secondary ion extraction vary over the area of the sample surface,in particular when spot-to-spot movements
over large distances (few mm) are required.
The sequence is as follows: |
![]() |
(4) Mass calibration
|
|
|
Prior to each analysis, the magnet is cycled over the entire mass range to be analyzed and centered for each mass
(acq_calibration_curve.vi). The mass range for each peak is set by two parameters: &bull Mass resolution (in "analysis_definition.vi" select "Anal. Par." and "Edit": type 5000 as mass resolution and confirm with "Set"). Note: if CIPS had to be restarted, the mass resolution is set to default = 300. &bull #dM. Default is 4.0. Once all peaks are centered, #dM can be changed to 1.3 (except for 204Pb which should remain at 4.0 or - if peak is close to the margins of the mass window - be increased to 6.0). Examples are shown for 204Pb, 206Pb, and 208Pb. Note that the 204Pb signal is typically too small to perform magnet centering on 204Pb directly. Instead, a nearby reference mass (94Zr2O) is used to center the magnet (blue cursor) whereas the B-filed during analysis is set for 204Pb (red cursor). &bull If peak is well defined, press "Center" + "Valid" &bull If intensities of 207Pb, 208Pb, or even 206Pb are low (<100 cps), be aware that there is a risk of miscentering the peak. This could also accidentally happen for the case a nearby intereference is higher in intensity (see Fig. for 208Pb). &bull If miscentering accidentally has occurred (note "crazy" B-field value in 208Pb scan), a window "mass is far from the calibration..." pops up after pressing "valid". Select "cancel", go back to a previous mass, then again to the miscentered mass. This time, find peak manually by dragging cursor. If insecure about peak position, the following is recommended: &bull Check mass centering on AS3 only; or &bull Manually drag the cursor relative to how the B-field values changed on 94Zr2O+ &bull After cycling through all mass stations, press "Apply" on calibration.vi window. If masses were off by >10 bits, redo centering: "Quit" calibration.vi and "Calibrate from condition" again. &bull "Apply" on "Analysis Definition" &bull "Acquire" window pops up: you are ready to press "Start". Change sample name! |
![]() |
(5)
Presputtering and data acquisition
|
|
| In order to stabilize the secondary ion signal and to remove surface contamination, the primary beam is focused on the sample surface for a certain time period prior to data acquisition. In this example (right), the duration of the presputtering was 3 min. Subsequently, intensities were measured in 15 magnet cycles. Isotopic ratios are then calculated using raw intensities corrected for electron multiplier dead-time and drift-corrected mid-analysis values. | ![]() |
(6) Interelement
calibration and precision
|
|
|
Interelement fractionation between U, Th and Pb during ion microprobe analysis is monitored using well-characterized standards of known isotopic composition (e.g., AS3, 91500). Resulting age uncertainties critically depend on how well this fractionation can be monitored. Typically, raw U-Pb ratios are corrected for common-Pb (e.g., by using the 204Pb signal). Interelement fractionation between U and Pb commonly varies between ~5 and ~10 (calculated as (Pb/U)measured/(Pb/U)true). These values are plotted as 206Pb+/U+ RSF (relative sensitivity factor) against UO+/U+ and regressed as a linear function (slopes of ~0.5±0.2 are typically determined). This allows for calibrating the U-Pb relative sensitivity as a function of UO+/U+ measured on the unknowns. |
![]() |
|
Literature: Schmitz MD, Bowring SA, and Ireland T (2003) Evaluation of Duluth Complex anorthositic series (AS3) zircon as a U-Pb geochronological standard: New high-precision isotope dilution thermal ionization mass spectrometry results. GCA 67, 3665-3672 Stern RA, and Amelin, Y. (2003) Assessment of errors in SIMS zircon U-Pb geochronology using a natural zircon standard and NIST SRM 610 glass. Chem Geol 197, 111 – 142 |